Tri-mode Image Intensifier /span>
- Dual High Resolution Monitors
Options:
- Vascular Application
- Tri-Mode 12"/9/6 II
Has a Rotating Anode
Specifications:
X-Ray System
Specialty Packages Available
• Extended processing
• Vascular functions
• Extended vascular functions
Generator
• 2.1 kW full-wave
• Up to 110 kVp
• Up to 20 mA for radiographic film exposures
• Continuous fluoro boost mode up to 7.3 mA
• Full power from outlet:
- 50 Hz or 60 Hz
- 100V-15A to 240V-10A
X-Ray Tube
• Fixed anode X-ray tube
• Focal spots: 0.6 mm and 1.4 mm
• Anode heat capacity: 50,100 HU
• Anode cooling rate: 20,200 HU/min
• Housing heat capacity: 1,200,000 HU
• Anode angle: 12°
Collimator
• Iris collimation with dual opposing lead shutters allowing elliptical, round and rectangular collimation
• Minimum beam size < 5 X 5 cm2
• Iris and shutters are continuously variable
• All functions remotely controlled from the C-Arm control panel
Fluoroscopy Mode
• Focal spot: 0.6 mm
• kVp range: 40 – 110 kVp
• mA range: 0.1 – 3.1 mA Low Dose fluoro mode
• mA range: 0.2 – 7.3 mA continuous High Definition fluoro boost mode
• Auto and manual modes
• Continuous or Pulsed operation
• APF varies mA, kVp and camera gain
• User specific APF tables
Pulsed Fluoroscopy Mode (optional)
• Focal spot: 0.6 mm
• kVp range: 40 – 120 kVp
• mA range: 0.4 – 12 mA
• Pulse rate: 6.25 or 12.5 pulses per second
• Pulse width: 11 to 37 milliseconds
• Computer controlled APF, mA, kVp and camera gain
Pulsed Fluorography Mode (optional)
• Focal spot: 0.6 mm
• kVp range: 40 – 110 kVp
• mA range: Up to 60 mA
• Pulse rate: 3 to 30 pulses per second (optional)
• Pulse width: 8.0 / 9.5 / 11.1 milliseconds
• Computer controlled camera iris, mA, kVp, & camera gain
Radiographic Mode
• Focal spot: 1.4 mm
• kVp range: 40 – 105 kVp
• Fixed 20 mA
Video Imaging System
Image Intensifier
9 in Image Intensifier
• Tri-mode 9 in / 7 in / 5 in image intensifier
- 9 in.: 1.6 lp/mm
- 7 in.: 2.2 lp/mm
- 5 in.: 2.9 lp/mm
• DQE: 65% (typical)
12 in Image Intensifier (optional)
• Tri-mode 12 in / 9 in / 7 in image intensifier
- 12 in.: 1.2 lp/mm
- 9 in.: 1.6 lp/mm
- 7 in.: 2.2 lp/mm
• DQE: 65% (typical)
Video Camera
• High resolution CCD camera
• 525 line/frame
• 360° motorized rotation
• Left-right image reversal
• Top-bottom image reversal
• Negative mode
• Computer controlled features: (optional)
- Gain
- Blanking
- Camera iris
Video Monitors
• Dual 17 in standard resolution, anti-glare, anti-static
• Ambient room light sensor
• Multi-LED X-ray "ON" indicator
Hardcopy Options
• Thermal printer (optional)
• DICOM (optional)
• Image Capture Device (optional)
C-Arm Dimensions
• Height: 69 / 89 in (1750 / 2250 mm)
• Width: 35 in (880 mm)
• Length: 72 in (1840 mm)
• Weight: 695 lbs 300 kg
Workstation Dimensions
• Height: 69 in (1760 mm)
• Width: 37 in (930 mm)
• Depth: 37 in (760 mm)
• Weight: 463 lbs 210 kg